A COMPARATIVE STUDY ON VACUUM AND ATMOSPHERIC PLASMA SPRAYED TANTALUM COATINGS FOR THE MODIFICATION OF TITANIUM IMPLANTS
Ding, Ding; Xie, Youtao; Li, Kai; Huang, Liping; Zheng, Xuebin
2019-10-01
Source PublicationSURFACE REVIEW AND LETTERS
ISSN0218-625X
Volume26Issue:8
SubtypeArticle
AbstractIn order to improve the mechanical compatibility and cytocompatibility of titanium implants, tantalum coatings were prepared using plasma spraying technology. Tantalum coatings have been deposited via atmospheric plasama spraying (APS) and vacuum plasma spraying (VPS) methods, and then their morphologies, porosities, bonding strengths and elastic modulous were investigated. In vitro cytocompatibility of the two coatings was evaluated via human bone marrow stromal cells (hBMSCs). The results show that oxidation phenomenon was observed for the APS tantalum coatings, while less oxidation was found in the VPS tantalum coatings. Compared with APS tantalum coatings, the VPS tantalum coatings have a more compact microstructure and less impurity content, resulting in a better bonding with the titanium substrate. VPS tantalum coatings were measured to have a lower elastic modulus and a higher hardness than APS Tantalum coatings. Electrochemical corrosion of the coatings were examined and the VPS tantalum coating showed improved chemical stability. Besides, bone marrow stem cells (BMSCs) adhere to and spread well on the surface of both VPS and APS tantalum coatings without significant difference. The proliferation rate of BMSCs is higher on VPS tantalum coating surface than on APS tantalum coatings. Our results suggest that VPS tantalum coatings are more suitable for the application of surface modification of titanium implant due to their lower elastic modulus and better chemical stability for higher mechanical compatibility and cytocompatibility.
KeywordTantalum coating vacuum plasma spraying elastic modulus cytocompatibility
DOI10.1142/S0218625X19500483
Language英语
WOS Research AreaChemistry ; Physics
PublisherWORLD SCIENTIFIC PUBL CO PTE LTD
Citation statistics
Document Type期刊论文
Identifierhttp://ir.sic.ac.cn/handle/331005/26799
Collection中国科学院上海硅酸盐研究所
Recommended Citation
GB/T 7714
Ding, Ding,Xie, Youtao,Li, Kai,et al. A COMPARATIVE STUDY ON VACUUM AND ATMOSPHERIC PLASMA SPRAYED TANTALUM COATINGS FOR THE MODIFICATION OF TITANIUM IMPLANTS[J]. SURFACE REVIEW AND LETTERS,2019,26(8).
APA Ding, Ding,Xie, Youtao,Li, Kai,Huang, Liping,&Zheng, Xuebin.(2019).A COMPARATIVE STUDY ON VACUUM AND ATMOSPHERIC PLASMA SPRAYED TANTALUM COATINGS FOR THE MODIFICATION OF TITANIUM IMPLANTS.SURFACE REVIEW AND LETTERS,26(8).
MLA Ding, Ding,et al."A COMPARATIVE STUDY ON VACUUM AND ATMOSPHERIC PLASMA SPRAYED TANTALUM COATINGS FOR THE MODIFICATION OF TITANIUM IMPLANTS".SURFACE REVIEW AND LETTERS 26.8(2019).
Files in This Item:
There are no files associated with this item.
Related Services
Recommend this item
Bookmark
Usage statistics
Export to Endnote
Google Scholar
Similar articles in Google Scholar
[Ding, Ding]'s Articles
[Xie, Youtao]'s Articles
[Li, Kai]'s Articles
Baidu academic
Similar articles in Baidu academic
[Ding, Ding]'s Articles
[Xie, Youtao]'s Articles
[Li, Kai]'s Articles
Bing Scholar
Similar articles in Bing Scholar
[Ding, Ding]'s Articles
[Xie, Youtao]'s Articles
[Li, Kai]'s Articles
Terms of Use
No data!
Social Bookmark/Share
All comments (0)
No comment.
 

Items in the repository are protected by copyright, with all rights reserved, unless otherwise indicated.