Application of K beta/K alpha in selecting calibration standards for X-ray fluorescence analysis
She, Xiaomeng; Zhu, Zhixiu; Gao, Jie; Qian, Rong; Sheng, Cheng; Shen, Ruxiang; Zhuo, Shangjun
2019-11-01
Source PublicationX-RAY SPECTROMETRY
ISSN0049-8246
Volume48Issue:6Pages:664
SubtypeArticle
AbstractA method for improving the accuracy in X-ray fluorescence spectroscopy analysis with pressed pellets was established through matching the calibration standards by the K beta/K alpha ratios. The K beta/K alpha ratios of four major elements (Al, Si, Ca, and Fe) were measured in 14 certified reference materials (CRMs) for geochemistry. The deviation of the biggest K beta/K alpha values relative to the smallest ones for the same element was from 1.52% (Si) to 20.98% (Al) when pressed pellets were used. The relative deviation of K beta/K alpha values for Ca was as high as 3.56% even for two soil CRMs (GBW07404 and GBW07408). X-ray diffraction analysis confirmed that the Ca elements in the two CRMs were in distinct phases, thereby leading to the different K beta/K alpha values. The Fundamental Parameter Method (FPM) results of two calibration strategies were compared. When one of the 14 CRMs was selected as unknown sample whereas the other 13 CRMs were used for calibration to determine Al2O3, SiO2, CaO, and Fe2O3, the relative errors were 10.80%, 8.77%, 8.33%, and 10.91%, respectively. Another strategy, where only one CRM was used for calibration of samples with large errors, was adopted by matching K beta/K alpha values. The results of calibrating using CRM with similar K beta/K alpha values were more accurate than those with different K beta/K alpha values.
DOI10.1002/xrs.3085
Language英语
WOS Research AreaSpectroscopy
PublisherWILEY
Citation statistics
Document Type期刊论文
Identifierhttp://ir.sic.ac.cn/handle/331005/26736
Collection中国科学院上海硅酸盐研究所
Recommended Citation
GB/T 7714
She, Xiaomeng,Zhu, Zhixiu,Gao, Jie,et al. Application of K beta/K alpha in selecting calibration standards for X-ray fluorescence analysis[J]. X-RAY SPECTROMETRY,2019,48(6):664.
APA She, Xiaomeng.,Zhu, Zhixiu.,Gao, Jie.,Qian, Rong.,Sheng, Cheng.,...&Zhuo, Shangjun.(2019).Application of K beta/K alpha in selecting calibration standards for X-ray fluorescence analysis.X-RAY SPECTROMETRY,48(6),664.
MLA She, Xiaomeng,et al."Application of K beta/K alpha in selecting calibration standards for X-ray fluorescence analysis".X-RAY SPECTROMETRY 48.6(2019):664.
Files in This Item:
There are no files associated with this item.
Related Services
Recommend this item
Bookmark
Usage statistics
Export to Endnote
Google Scholar
Similar articles in Google Scholar
[She, Xiaomeng]'s Articles
[Zhu, Zhixiu]'s Articles
[Gao, Jie]'s Articles
Baidu academic
Similar articles in Baidu academic
[She, Xiaomeng]'s Articles
[Zhu, Zhixiu]'s Articles
[Gao, Jie]'s Articles
Bing Scholar
Similar articles in Bing Scholar
[She, Xiaomeng]'s Articles
[Zhu, Zhixiu]'s Articles
[Gao, Jie]'s Articles
Terms of Use
No data!
Social Bookmark/Share
All comments (0)
No comment.
 

Items in the repository are protected by copyright, with all rights reserved, unless otherwise indicated.